Abstract

    Open Access Research Article Article ID: GJODMS-2-107

    Personalized Indirect Calorimeter for Energy Expenditure (EE) Measurement

    Francis Tsow*, Erica Forzani*, Nongjian Tao**, Xiaojun Xian, Ashley Quach, Devon Bridgeman

    Background and aims: A personal indirect calorimeter allows everyone to assess resting and non-resting energy expenditure, thus enabling accurate determination of a person’s total calorie need for weight management and fitness. The aim of this study is to compare the performance of a new personal metabolic rate tracker based on indirect calorimetry, Breezing®, with the Douglas bag method, the gold standard method for energy expenditure (EE) measurement.

    Methods: Energy expenditures (EE) at rest and during activities, and respiratory quotient (RQ) were measured for 12 healthy subjects, including 7 male and 5 female under different living conditions. A total of 314 measurements were performed with Breezing®, and the results were compared with those by the Douglas bag method.

    Results: R-squared correlation coefficients (R2) between the data obtained with Breezing® and the Douglas bag method were 0.9976, 0.9986, 0.9981, and 0.9980, for VO2, VCO2, EE, and RQ respectively.

    Conclusions: The EE and RQ values determined by Breezing® are in good agreement with those by the Douglas bag method.

    Keywords: Indirect calorimetry; Metabolism tracker; Energy expenditure; Respiratory quotient

    Published on: Mar 29, 2015 Pages: 4-8

    Full Text PDF Full Text HTML DOI: 10.17352/2455-8583.000007
    CrossMark Publons Harvard Library HOLLIS Search IT Semantic Scholar Get Citation Base Search Scilit OAI-PMH ResearchGate Academic Microsoft GrowKudos Universite de Paris UW Libraries SJSU King Library SJSU King Library NUS Library McGill DET KGL BIBLiOTEK JCU Discovery Universidad De Lima WorldCat VU on WorldCat

    Indexing/Archiving

    Pinterest on GJODMS